Urisys 1100
®
Urine Analyzer
Policies
and
Procedures
for use with Urisys 1100
®
Urine Analyzer
This is a CLIA-Waived System
Roche Diagnostics is not responsible for any changes made to this diskette by the customer.
Introduction
Roche Diagnostics is pleased to assist you in developing policies and procedures and implementing regulations of the
Clinical Laboratory Improvement Amendment of 1988 (CLIA ‘88) in your facility. This guide contains sample policies
and procedures for use with the Urisys 1100
®
Urine Analyzer. These sample policies and procedures are intended only
as a guide, and it is important that you make whatever adjustments may be required as a result of a change in policy or
a change in law or regulations or their interpretation.
A total quality management program should include written policies and testing procedures that are customized for
your particular facility. Any policies and procedures that involve using Chemstrip urine test strips should be written
prior to the start of any urine testing.
This guide provides sample policies and procedures that meet the Section 493.1211 requirements. Blank spaces let you
customize the policies and procedures with information specific to your facility. You are encouraged to highlight impor-
tant areas and procedures currently in use.
We suggest you store a Urisys 1100
®
Urine Analyzer Operator’s Manual, Chemstrip urine test strip package insert and
control package insert in the Product Information section for quick reference. You may also want to include the following
clinical guidelines and regulations in the Regulatory Requirements section.
• JCAHOGuidelines
• CAPGuidelines
• CLIA‘88Guidelines
• StateGuidelines
• BloodbornePathogenStandard
If you have any questions, please call Roche Diagnostics Technical Service Center at 1-800-428-4674, available 7 days
per week, 24 hours a day, and 365 days per year.
Approval Process
The following policies and procedures must be approved, signed and dated by the laboratory director. If a new director
is placed in charge, the procedures must be re-approved, signed, and dated by the new director. In addition, any
changes to the manual must be approved, signed, and dated by the laboratory director.
You must maintain a copy of each procedure with the date of the initial use and the date the test was discontinued.
Keep these records for at least two years after the procedure has been discontinued.
Approval date ____________________ Approved by: _____________________________________________________
Reviewed/Revised
Date By Whom Title
_____________________ _____________________________________ ___________________________________
_____________________ _____________________________________ ___________________________________
_____________________ _____________________________________ ___________________________________
_____________________ _____________________________________ ___________________________________
_____________________ _____________________________________ ___________________________________
_____________________ _____________________________________ ___________________________________
_____________________ _____________________________________ ___________________________________
_____________________ _____________________________________ ___________________________________
Purpose
To establish safe, accurate urine test results to be used for screening and monitoring patient status.
General Policies and Procedures
• ThepoliciesandprocedurespertainingtourinalysiswiththeUrisys1100
®
Urine Analyzer are reviewed at least once
a year.
• Anychangestothetextastheyapplytomethodologyortestingpolicyaresignedbytheappropriateauthority,
dated, and inserted in the policies and procedures manual.
• Obsoleteorsupersededproceduresarereplacedinthemanual,butareretainedforcompliancepurposes.
• ThemostrecentpackageinsertsandtheUrisys 1100
®
Urine Analyzer Operator’s Manual provided by Roche
Diagnostics should be referenced for the most recent procedural information.
Table of Contents
I. PoliciesandProcedures ........................................................................................................................................................................... 1
A. PrincipleofOperation .................................................................................................................................................................. 1
B. Specimen Collection ..................................................................................................................................................................... 3
C. Test Strip Storage and Stability ................................................................................................................................................ 4
D. Calibration of the Urisys 1100
®
Urine Analyzer .................................................................................................................. 5
E. Testing with the Urisys 1100
®
Urine Analyzer ..................................................................................................................... 7
F. Documentation of Urinalysis Results ..................................................................................................................................... 11
G. ExpectedValues ............................................................................................................................................................................. 12
H. QualityControlProcedure .......................................................................................................................................................... 13
I. DocumentationofQualityControlResults .......................................................................................................................... 15
J. LimitationsoftheProcedure ..................................................................................................................................................... 16
II. PreventiveMaintenance .......................................................................................................................................................................... 18
III. InfectionControlGuidelines ................................................................................................................................................................... 20
IV. OperatorCertification/Recertification ................................................................................................................................................ 21
V. Appendix........................................................................................................................................................................................................ 25
VI. LogSheets .................................................................................................................................................................................................... 28
VII. RegulatoryRequirement .......................................................................................................................................................................... 44
VIII. ProductInformation ................................................................................................................................................................................... 45
Urisys 1100
®
andCHEMSTRIParetrademarksofaMemberoftheRocheGroup.
All other trademarks are the property of their respective holders.
©2003 Roche Diagnostics. All rights reserved. REF 03755541001 (03)
1
*HospitalsmayuseChemstrip10UAUrineTestStrips.
I. Policies and Procedures
A. Principle of Operation
The Roche Diagnostics Urisys 1100
®
Urine Analyzer is a semi-automated analyzer intended for in vitro semi-quantitative
determinationofurineanalytes.TheanalyzerisdesignedtoreadandevaluatetheresultsofChemstrip10MD*,
Chemstrip7,andChemstrip5OBUrineTestStrips.Theseurineteststripsaremultiparameterstripsusedforthedeter-
minationofspecificgravity,pH,leukocytes,nitrite,protein,glucose,ketones,urobilinogen,bilirubinandbloodinurine.
The Urisys 1100
®
Urine Analyzer is a reflectance photometer. It reads the urine test strips under standardized condi-
tions, saves the results to memory and outputs them via its own built-in printer and/or serial interface.
The Urisys 1100
®
Urine Analyzer standardizes urine test strip results by eliminating factors known to affect visual evalu-
ation of urine test strips, such as; variable lighting conditions at the workplace, individual skill levels at matching test
strip pad colors, failure to keep the prescribed testing times, and clerical errors.
The test strip is placed on a sliding tray, and a stepping motor moves it under the reading head, which remains station-
ary. The analyzer reads the reference pad, followed by each of the test pads on the strip. The reading head contains
LEDs that emit light at various wavelengths. Reading is done electro-optically as follows:
The LED emits light of a defined wavelength onto the surface of the test pad at an optimum angle. The light hitting the
test zone is reflected proportionally to the color produced on the test pad and is picked up by the detector, a phototran-
sistor positioned directly above the test zone. The phototransistor sends an analog electrical signal to an A/D converter,
which changes it to digital form. The microprocessor then converts this digital reading to a relative reflectance value by
referring it to a calibration standard.
Finally, the system compares the reflectance value with the defined range limits (reflectance values which are pro-
grammed into the analyzer for each parameter) and outputs a semi-quantitative result.
Each test pad is read photometrically after about 55-65 seconds. In strongly alkaline urine samples, the Urisys 1100
®
Urine Analyzer automatically corrects the result of the specific gravity test.
Thefollowingisalistoftestprinciplesspecifictoeachparameter.PleasereferencethemostrecentChemstrip urine
test strip package insert for modifications in test principles.
• Specific Gravity:Inthepresenceofcations,protonsarereleasedbyacomplexingagentinthetestpad.Theindi-
cator bromthymol blue changes from blue via blue-green to yellow.
• pH Test: The test pad contains the indicators methyl red and bromthymol blue. These indicators give clearly distin-
guishablecolorsoverthepHrangeof5to9.
1, 2
Colors range from orange through yellow and green to blue.
• Leukocyte Test:Granulocyticleukocytescontainesteraseswhichcatalyzethehydrolysisofanindoxylcarbonic
acidestertoindoxyl.Theindoxylformedreactswithadiazoniumsalttoproduceapurplecolor.
• Nitrite Test: Nitrite, if present, reacts with an aromatic amine to give a diazonium salt, which, by coupling with a
further compound, yields a red-violet azo dye.
3, 4, 5
• Protein Test: The test is based on the color change of the indicator 3’, 3”, 5’, 5”-tetrachlorophenol-3, 4, 5, 6-tetrabro-
mosulfophthalein in the presence of protein. A positive reaction is indicated by a color change from yellow to light
green/green.
6, 7, 8
2
• Glucose Test:Glucosedetectionisbasedontheenzymaticglucoseoxidase/peroxidase(GOD/POD)method.The
reactionutilizestheenzymeglucoseoxidasetocatalyzetheformationofgluconicacidandhydrogenperoxidefrom
theoxidationofglucose.Inturn,asecondenzyme,peroxidase,catalyzesthereactionofhydrogenperoxidewiththe
chromogentetramethylbenzidinetoformagreendyecomplex.Apositivereactionisindicatedbyacolorchange
from yellow to green.
9, 10
• Ketone Test: Based on the principle of Legal’s Test, sodium nitroprusside and glycine react with acetoacetate and
acetoneinanalkalinemediumtoformavioletdyecomplex.Apositiveresultisindicatedbyacolorchangefrom
beige to violet.
11, 12
• Urobilinogen Test:Urobilinogeniscoupledwith4-methoxybenzene-diazonium-tetrafluoroborateinanacid
medium to form a red azo dye.
13
• Bilirubin Test: The detection of bilirubin is based on the coupling reaction of a diazonium salt with bilirubin in an
acid medium. The reaction yields a pink to red-violet color proportional to the total bilirubin concentration.
14
(Some
users may describe this as a cream to peach color.)
• Blood Test:Hemoglobinandmyoglobin,ifpresent,catalyzetheoxidationoftheindicatorbytheorganicperoxide
contained in the test pad. Intact erythrocytes hemolyze on the test pad and liberate hemoglobin, which produces a
green dot. Since the test pad absorbs several microliters of urine, more erythrocytes become visible than would cor-
respond to 1µL.
15-19
Separate sets of color blocks are representative of erythrocytes and hemoglobin. Scattered or
compacted green dots on the yellow test pad are indicative of intact erythrocytes, or myoglobin.
• Compensation Pad: The compensation area is not impregnated with reagent. This pad allows the instrument to
compensate for the intrinsic color of urine, which may affect test results.
3
B. Specimen Collection
Policy
• Useafreshlyvoidedurinespecimen,first-morningurinespecimen,orpost-prandialurinespecimen.
• Whentestingurinespecimens,everyprecautionmustbetakentoavoidcontaminationofthespecimenandthe
operator.
• Becauseofthepotentiallyhazardousnatureofhandlingurineandthepossibilityofitscontaminationwithblood,it
isrecommendedthatdisposablelatexglovesandlaboratorycoatsbeusedwhencollectingspecimens.Consultyour
facility’s infection control policy.
• Instructionsforcollectionofaclean-catchmidstreamurinespecimenmaybefoundinSection V. Appendix.
Procedure
1. Follow physician’s order(s).
2. Assemble appropriate supplies.
• Cleansingtowelette
• Water-saturatedsponge
• Cleanurinecontainerwithlid
3. Instruct patient to wash genital area (do not use towelettes with strong cleansing agents), rinse with water, and then
obtain a clean-catch midstream urine sample in the clean container. Refer to Section V. Appendix for instructions.
4. Test urine as soon as possible after collection to avoid deterioration of specimen. If specimen cannot be tested
within two hours of collection, refrigerate immediately (at 2°-8°C) in a closed container. Bring specimen to room
temperaturebeforetesting.Mixthoroughlybeforetesting.
5. Ifurineisnottestedimmediately,mixspecimenthoroughlybycappingcontainerandswirlingseveraltimesbefore
testing to avoid incorrect results. Do not centrifuge or add preservatives to the urine specimen.
4
C. Test Strip Storage and Stability
Policy
• StoreChemstripurineteststripsat2°C(36°F)–30°C(86°F).Donotfreeze.
• OpenedChemstripurineteststripsarestableuntiltheexpirationdateontheviallabelwhenstoredintheoriginal
capped vial. The vial must be closed immediately after use, using the original cap.
5
*HospitalsmayuseChemstrip10UAUrineTestStrips.
D. Calibration of the Urisys 1100
®
Urine Analyzer
(withChemstrip10MD*TestStrips)
Policy
• CalibrationisrequiredifusingtheChemstrip10MD*TestStrips.(WhenusingChemstrip5OBorChemstrip7Test
Strips, the calibration procedure is not required. These test strips are calibrated using the gray reference pad on the
test strip tray.)
• TheUrisys1100
®
Urine Analyzer must be calibrated every seven days or when indicated by the analyzer.
• TheChemstripUrineAnalyzerCalibrationStrips(CatalogNumber418007)shouldremaininthevialuntiljustbefore
use. Do not touch the elevated gray areas on the strips.
• Thecalibrationstripsshouldbeusedonlyonce.
• Ifyoudonotobtainasuccessfulcalibration,theanalyzerwillnotprocesssamples.
• CalibrationshouldbedocumentedonthePreventive Maintenance Log.
Procedure
1. Plugthepowercordprovidedwiththeanalyzerintothepoweroutletattherearoftheanalyzerandthenintoawall
outlet. Turn the analyzer on. The analyzer will then automatically perform a self check. If the self check is completed
successfully,themessage“selfcheckOK”printsalongwiththetimeanddate.
2. Makesuretheteststriptrayiscleananddry.
Note: After a successful self check, the operator may have to enter an operator ID and password (if these options have
been selected).
3. Ifthemessage“REPEATCALIBRATION”isinthedisplay,pressSTART.IftheanalyzerisinReady-to-Measuremode,
presstheleftfunctionkeytoselect“Calibr”.Themessage,“STARTCALIBRATION”appearsonthedisplay.
4. Remove a calibration strip from its container, being careful not to touch the pads. Do not allow the calibration strip
to come into contact with urine.
5. With the retaining bar open, place the calibration strip, pads facing upward, onto the test strip tray and insert the
front edge of the strip under the plastic clip.
6. PresstheSTARTbutton.Youwillhearabeepandthenthetrayadvancesslightly,theretainingbarcloses,andthe
calibration strip is read. When calibration is complete, the tray moves to the start position and the retaining bar
opens.
7. Remove the calibration strip and discard. Remember that each calibration strip is used only once.
8. If the calibration is valid, results are stored in memory, and a report with the reflectance values, time, and date auto-
matically prints. Reflectance values for positions 1 through 11 for the orange LED are printed in the middle column
and for the green LED in the right column.
6
9. If the results for the reference pad or the calibration strip are outside the programmed tolerances, one of the follow-
ing messages appears on the display:
 • REFERENCEPADERROR
• “CALIBRATIONINVALID”or
• “CALIBRATIONERROR”
If a calibration error occurs, repeat the calibration procedure using a new calibration strip.
1. PresstheSTARTbuttonandthedisplaywillreturntothe“STARTCALIBRATION”menu.
2. PlacethenewcalibrationstripontheteststriptrayandpresstheSTARTbuttonagain.
3. Ifthecalibrationisvalid,themessage“CALIBRATIONOK”prints.
4. Ifthecalibrationfailsasecondtimeandacalibrationerrorappearsonthedisplay,refertotheOperator’sManualor
call Roche Diagnostics Technical Service at 1-800-428-4674, 7 days a week, 24 hours a day, and 365 days per year.
7
E. Testing with the Urisys 1100
®
Urine Analyzer
Policy
• OnlyacertifiedoperatormayperformaurinetestwiththeUrisys1100
®
Urine Analyzer.
• Becauseofthepotentiallyhazardousnatureofhandlingurineandthepossibilityofitscontaminationwithblood,it
isrecommendedthatdisposablelatexglovesandlaboratorycoatsbeusedwhencollectingspecimensandperform-
ing test procedures. Consult your facility’s infection control policy.
• UrineshouldbehandledatBiosafetyLevel2asrecommendedforanypotentiallyinfectiousmaterialintheCenters
forDiseaseControl/NationalInstitutesofHealthmanual,Biosafety in Microbiological and Biomedical Laboratories,
1988.
24
Universal precautions may apply if the urine is contaminated with blood or if required by the infection control
policy of the facility.
• OutdatedChemstripurineteststripsarediscarded.
Procedure
1. The following equipment is assembled prior to testing:
Urisys 1100
®
Urine Analyzer
Chemstrip urine test strips
Patientspecimen
Disposablelatexgloves(ifrequiredbyfacility)
Laboratory coat (if required by facility)
Absorbent paper
2. ChecktheexpirationdateontheChemstripurineteststripvialpriortotesting.
3.Plugthepowercordprovidedwiththeanalyzerintothepoweroutletattherearoftheanalyzerandthenintoawall
outlet. Turn the analyzer on. The analyzer will then automatically perform a self check. If the self check is completed
successfully,themessage“selfcheckOK”printsalongwiththetimeanddate.
Note: After successful self check, the operator may have to enter their operator ID and password if these options have
been activated.
4. The Urisys 1100
®
Urine Analyzer is ready for measurement as long as valid calibration values are stored in the ana-
lyzer. Refer to Section I.D. Calibration of the Urisys 1100
®
Urine Analyzer of this manual.
5. Verifytheteststriptrayiscleanofanyresidue.Ifthereisresidueonthetray,refertoSection II. Preventive Mainte-
nance of this manual for the cleaning procedure. Urine residue on the test strip tray may result in inaccurate results.
6.Ifthesamplehasbeenrefrigerated,allowittocometoroomtemperature.Mixthespecimenbycappingthecon-
tainer and swirling several times if not freshly voided.
8
7. Remove the cap from the test strip vial and remove one Chemstrip urine test strip.
8. Dip the test strip for one second in the urine sample. Draw the edge of the test strip along the rim of the specimen
container.Gentlytouch(onesecond)thelongedgeoftheteststriptoapieceofabsorbentpapertoremoveany
excessurine.
9. IftheanalyzerisinStandbymode,pressSTARTtoreturntoReady-to-Measuremode.Theteststriptrayandthe
retainingbarmustbeintheopenposition.Placetheteststrip,withthepadsfacingupward,ontotheteststriptray
and insert the front edge of the strip under the plastic clip. It is important that the strip is correctly positioned and
the START button is pressed within 5-10 seconds of dipping the strip.
10.PresstheSTARTbutton.Youwillhearabeepandthenthetrayadvancesslightly,andtheretainingbarcloses.First
the gray reference pad located on the test strip tray is read.
11. If the test strip is not lined up straight on the test strip tray, gently move it until it is properly lined up under the
retaining bar. Be careful not to move the tray.
12.Approximately60secondsaftertheSTARTkeyispressed,allteststrippadswillberead.Theresultswillprintand
thenextsamplenumberappearsonthedisplay.
13. The test strip tray will return to the start position and the retaining bar will open. Remove the used test strip and
wipe the test strip tray as needed with a lint-free cloth to remove any urine residue. Refer to Section II. Preventive
Maintenance of this manual for the cleaning procedure.
14.Properlydisposeoftheurine,urinecontainerandteststrip.
Guidelines
• Mechanicalproblemsareevidentwhentheanalyzerdisplaysanerrormessage.Achemistryproblemmaybecome
evidentwithanunexpectedresult.Abnormalresultsareflaggedwithanasterisk.Ifyouobtainanunexpectedresult
on any chemistry, refer to the Limitations and Performance Characteristics sections of the Chemstrip urine test strip
package insert.
• Ifyouarecomparingtheanalyzervaluesagainstthevisualstripvaluesandthedeviationismorethanoneconcen-
tration block, one of the following may have occurred:
• Teststripisnotpositionedcorrectlyintheteststriptray
• Powersupplyfailureorhighfrequencyinterferencebyotherdevices
• Visualstripisdeteriorated
• Impropertestingtechnique
If one of the above situations has occurred, repeat the measurement with a new test strip, being very careful to properly
positiontheteststrip.Removeanyexternalsourceofinterferences.
9
• Thefollowingisalistoferrormessagesandsuggestionsonhowtoeliminatetheproblem:
Error Code Error Description Cause Action
E1 Reference pad error Grayreferencepadontheteststriptray
soiled or damaged.
Switch off the instrument. If not visibly damaged, clean and
dry the pad. Insert the tray and wait for the self-check to
finish. If the error message reoccurs, call Roche Diagnostics
Technical Service Center.
E2 Wrong strip The test strip used is different from the
one for which the analyzer has been
programmed for (either Chem 10, Chem
5,orChem7).SeeOperator’sManual.
PresstheSTARTbutton.Repeatthemeasurementusingthe
test strip that the instrument has been programmed for.
E3 Strip measurement a) No test strip present on the tray.
b) Test strip is incorrectly positioned on
the tray.
c) Urine on the test strip has dried.
d) Test strip has not been dipped in
urine.
PresstheSTARTbutton.Repeatthemeasurementwitha
newteststrip.Makesurethestripisinsertedcorrectlyand
that the retaining bar properly closes after the START but-
ton is pressed.
E4 Calibration Error Calibration values differ from those
obtained in the last valid calibration.
PresstheSTARTbutton.Repeatthecalibrationusinganew
calibrationstrip.Makesurethecalibrationstripisproperly
positioned under the clip on the test strip tray.
E5 Calibration Invalid Calibration values out of tolerance. Check the reference pad for soil or damage. Clean the ref-
erence pad or use the spare tray. Repeat the calibration
with a new calibration test strip. If the error reappears, call
Roche Diagnostics Technical Service Center.
E6 Chip Error The chip module underneath the printer
cover is missing, not making contact, or
is defective.
Switch off the analyzer, insert the chip module and switch
the instrument on again. If the error message appears
again, call Roche Diagnostics Technical Service Center.
E7 MissingTray Test strip tray is missing or not inserted
far enough to engage the motor.
Inserttheteststriptraycorrectly.(SeeSection4ofOpera-
tor’sManual.)PresstheSTARTbutton.
E8 TrayPositionError The positioning hole in the tray is dirty
or still wet after cleaning; the retaining
bar is open while the tray is advancing;
or the retaining bar mechanism is con-
taminated with urine and blocked.
Clean, blow through, or dry the positioning hole on the test
strip tray using a lint-free cloth. Remove any urinary depos-
its on both the top and bottom of the test strip tray. Insert
the tray again and press the START button. If the message
error persists, call Roche Diagnostics Technical Service
Center.
E9 Wrong tray The test strip tray is not the one pro-
grammed for the analyzer or the gray
reference pad is missing from the tray.
Grayreferencepadisscratchedordirty.
PresstheSTARTbutton.Thestriptypemenuisdisplayed.
Thestriptypemustmatchthetraytype.(SeeOperator’s
Manual,Section4.2.)Usethecorrecttrayandmakesure
thegrayreferencepadisclean.PresstheSTARTbutton.
E10 Light Barrier Error The light barrier used to control the
position of the test strip is defective or
the tray transport is blocked.
Pulloutthetrayandreturnittothestartposition.Pressthe
START button. If the error message persists, call Roche
Diagnostics Technical Service Center.
E11 MotorStepError The stepping motor drive is out of toler-
ance or the advance of the test strip
tray is blocked.
Carefully clean the tray. Clean and dry the positioning hole
using a lint-free cloth. Remove any urinary deposits on both
the top and bottom of the test strip tray. Insert the tray
again and press the START button. If the message error
persists, call Roche Diagnostics Technical Service Center.
E12 OpticsError The reference pad is missing from the
test strip tray or an LED or phototrans-
istor is defective.
Attachthereferencepad.PresstheSTARTbutton.Ifthe
error message appears again, call Roche Diagnostics Tech-
nical Service Center.
E13 PrinterError Printerconnectionfaultorpaperjam.
This message will not appear if there is
no paper.
Switch off the analyzer. If there is a paper jam, clear it.
Switch the analyzer on again. If the error message appears
again, call Roche Diagnostics Technical Service Center.
10
Error Code Error Description Cause Action
E14 Interface Error UnabletotransferdatatoPCorhostin
bi-directional mode.
Checkthedatacable.VerifythatthePCorhostisreadyto
receive data. Use the “Send” function to transfer data or
press“Home”toresumeReady-to-Measurestatus.
E15 REFER-
ENCEPAD
ERROR
BOTTOM!
The bottom portion of the reference pad
on the tray is soiled or damaged.
see E1
E16 REFER-
ENCEPAD
ERRORTOP!
The top portion of the reference pad on
the tray is soiled or damaged.
see E1
E17INVALID
PASSWORD!
The entered password doesn’t match. Enter an correct password.
E18INVALID
OP.ID!
TheenteredOperatorIDisnotvalid. EnteravalidOperatorID.
E19 LIST
DOWNLOAD
FAILED
ThenewOperatorIDlistdownload
failed.
Noaction.After2secondsnextstatestartswitholdlistif
there was.
E20NO
VALIDLIST!
There is not a valid list at all in device. TrytodownloadanOperatorIDlistfromthehostorcon-
tinue without authenticated operator.
CheckMea-
surement
Analyzer prints out
software and chip
version number and
3-digit numbers
without naming the
parameters
Service function is activated. Pressthe“Back”functionkeytoreturntothemainmenu.
Valuesobtaineddo
not compare with
those from visual
examination.
a) Test strip incorrectly positioned on
the test strip tray.
b) Uncharacteristic test pad colors.
c) The wrong test strip, such as the cali-
bration strip, was used.
d) Electromagnetic interference from
other devices.
e) Improper storage or technique.
Repeat the measurement with a new test strip. Follow the
procedure carefully and make sure the test strip is inserted
properly.Repeatcalibration,ifnecessary.Movetheanalyzer
toanotherlocationifexternalinterferenceissuspected.
No printout “Printer:Off”isselectedortheprinter/
software is defective.
Choose“Printer:On”toreactivatetheprinter.Requesta
patientreportthroughthe“Print”function.Ifthisfails,acti-
vate the “Linefeed” function. If there is still no response, call
ROCHEDIAGNOSTICSTechnicalServiceCenter.
The analyzer will not
read, even though
the sequence num-
ber is displayed.
Switch the analyzer off and back on again.
11
F. Documentation of Urinalysis Results
Policy
• Thedate,time,initialsoftheoperator,patientname/patientIDnumber,andparametervaluesarerecordedonthe
patient chart or the appropriate log for your facility.
• Testrequisitions,testauthorizations,andtestresultsareretainedforaminimumoftwoyears.
• Testrequisitionorauthorizationincludes:
Patient’sname
Name and address (or other identifier) of the person requesting the test
Test to be performed
Date the sample was obtained
Date the test was performed
• Anaudittrailexists,linkingthepatienttestwiththeChemstripteststriplotnumberandcontrollotnumberused.
Procedure
1. Record the date, time, initials of the tester, patient name/patient ID number, and test values on the Urinalysis Report
Form of this guide or the appropriate log for your facility.
2. Indicate the type of specimen (first morning, random, clean-catch midstream, etc.) on the appropriate log.
(See Section VI. Log Sheets for sample log sheets.)
12
G. Expected Values
Guidelines
PleaserefertothemostrecentChemstrip urine test strip package insertforanychangestotheexpectedvaluesinfor-
mation.
• Specific Gravity: Random urine specimens vary from 1.001 to 1.035. Twenty-four hour urine specimens from nor-
mal adults with normal diets and fluid intake will have a specific gravity of 1.016 to 1.022.
22
• pH:UrinepHvaluesgenerallyrangefrom5to9.ThemostfrequentpHvaluesforfirst-morningspecimensin
healthy subjects are between 5 and 6.
• Leukocytes: Normal urine should produce no color reaction. “Trace” indicates a possible borderline situation, and
thetestshouldberepeatedonafreshurinesamplefromthepatient.Positiveandrepeatedtracefindingsindicate
the need for further testing of the patient and/or urine sample in accordance with the medically accepted proce-
dures for pyuria.
• Nitrite: A nitrite concentration as low as 0.05 mg/dL will produce a slightly pink coloration of the test pad. This indi-
cates a positive result.
• Protein: A color change from yellow to light green/green will occur if protein is present in urine. The concentrations
givenontheviallabelcorrespondwiththealbuminconcentrationinurine.Pathologicalproteinuriausuallywillpro-
duce persistent values above 30 mg/dL. Clinical significance of the trace result should be determined by additional
testing.
• Glucose: Due to the test’s sensitivity, glucose should not be detectable in normal urine. Therefore, any positive
reaction should be followed by further diagnostic evaluation of the patient, such as a quantitative blood glucose or a
glucose tolerance test.
• Ketones: Ketone bodies should not detected in normal urine with this test. Fasting or starvation diets may cause
positive results. In known pathological conditions such as diabetes, the presence of ketones may be useful as an
indexofmetabolicstatus.
• Urobilinogen:Concentrationsareusuallygreaterintheafternoonthanduringtheremainderoftheday.Valuesup
to 1 mg/dL are usually considered normal.
13
• Bilirubin:Innormalurine,bilirubinshouldnotbedetectablewiththistest.However,thetestisverysensitivetobili-
rubin (0.5 mg/dL will produce positive results), and any positive reaction indicates that further diagnostic evaluation
of the patient is needed.
• Blood:Erythrocyteexcretionupto5Ery/µLmaybeexpectedinnormalurine.
17, 18
Levels above this may require fur-
ther evaluation of the patient.
13
H. Quality Control Procedure
Policy
• Qualitycontroltestingwillbeperformedwithaminimumoftwo(2)qualitycontrolsolutionsaccordingtoyourfacili-
ty’s policy.
• Itisrequiredtorunapositiveandnegativecontroldailyandwhenanewvialofteststripsisopened(including
everylotchange).Valuesobtainedforthesecontrolsshouldfallwithinthelimitsestablishedbythelaboratoryorthe
control manufacturer.
• Checktheappropriateboxesforyourfacility’sadditionalqualitycontrolfrequency,beingsuretomeetallstandards
of your regulatory agencies.
q Each ________ (when? day, shift, etc.).
q Whentheteststripshavebeenexposedtoextremeheat,humidityorcold.
q When test results contradict clinical symptoms.
q As a way of checking operator competency.
q Other______________________________________________________________________________________.
• Ifaqualitycontroltestresultfallswithintheacceptablecontrolrange,itisacceptabletoproceedwithpatienttesting.
• Ifaqualitycontroltestresultfallsoutsideoftheacceptablecontrolrange,repeatthetestwithafreshChemstrip
urine test strip. Check the stability of the control being used. The problem must be corrected before proceeding with
patient testing.
• Anyqualitycontrolresultthatfallsoutsideoftheacceptablecontrolrange,alongwithanycorrectiveactionto
restore that result to acceptable range, must be recorded in the Urinalysis Quality Control Log or other quality control
log.
• Thecontrolsshouldbestoredatthepropertemperaturesaccordingtothemanufacturer’srecommendations.
• Anyoutdatedteststripsorcontrolswillbediscarded.
• TheurineteststriplotnumberandexpirationdatearefoundonthelabelofeachvialofChemstripurineteststrips.
This information should be recorded in the Urinalysis Quality Control Log or other quality control log.
• Chemstripurineteststripsmustbestoredattemperaturesbelow30°C(86°F).Donotfreeze.Teststripsarestable
untiltheexpirationdate.Teststripsmustbestoredinthesamecappedvialinwhichtheywerepackaged,andthe
vial cap must be immediately replaced after removal of a test strip.
• UrinecontrolsrecommendedforusewiththeUrisys1100
®
UrineAnalyzerareKOVA-Trol
®
andKOVA
®
Liqua-Trol™
(HYCORBiomedicalInc.),theDipperUrineDipstickControl(Quantimetrix)andLiquichek™(BIORAD).Referto
Section V. Appendix for more information about these controls.
• Insertyourfacility’squalitycontrolpreparationinformationafterthispage.
14
Procedure
1. The following equipment is needed for quality control testing:
Urisys 1100
®
Urine Analyzer
Chemstrip urine test strips
Commercially prepared urinalysis control
Disposablelatexgloves(ifrequiredbyfacility)
Laboratory coat (if required by facility)
Absorbent paper
2. Preparecontrolsaccordingtothemanufacturer’sinstructions.
3. ChecktheexpirationdateontheChemstripurineteststripvialpriortotesting.
4. The Urisys 1100
®
Urine Analyzer is ready for measurement as long as valid calibration values are stored in the ana-
lyzer. Refer to Section I.D. Calibration of the Urisys 1100
®
Urine Analyzer of this manual.
5. Verifytheteststriptrayiscleanofanyresidue.Ifthereisresidueonthetray,refertoSection II. Preventive Mainte-
nance of this manual for the cleaning procedure. Any residue on the test strip tray may result in inaccurate results.
6. Remove the cap from the test strip vial and remove one Chemstrip urine test strip.
7. Diptheteststripforonesecondinthecontrol.Drawtheedgeoftheteststripalongtherimofthecontainer.Gently
touch(onesecond)thelongedgeoftheteststriptoapieceofabsorbentpapertoremoveanyexcesscontrol.
8. IftheanalyzerisinStandbymode,pressSTARTtoreturntoReady-to-Measuremode.Theteststriptrayandthe
retainingbarmustbeintheopenposition.Placetheteststrip,withthepadsfacingupward,ontotheteststriptray
and insert the front edge of the strip under the plastic clip.
9. PresstheSTARTbutton.Youwillhearabeepandthenthetrayadvancesslightlyandtheretainingbarcloses.First
the gray reference pad located on the test strip tray is read.
10. If the test strip is not lined up straight on the test strip tray, gently move it until it is properly lined up under the
retaining bar. Be careful not to move the tray.
11.Approximately60secondsaftertheSTARTkeyispressed,allteststrippadswillberead.Theresultswillprintand
thenextsamplenumberappearsonthedisplay.
12. The test strip tray will return to the start position and the retaining bar will open. Remove the used test strip and
wipe the test strip tray as needed with a lint-free cloth to remove any residue. Refer to Section II. Preventive Mainte-
nance of this manual for the cleaning procedure.
13.Properlydisposeofthecontrolsolutionandteststrip.
15
I. Documentation of Quality Control Results
Policy
• Qualitycontrolrecordsareretainedforaminimumoftwoyears.
• Thedate,time,initialsoftheoperator,andqualitycontrolresultarerecordedintheUrinalysis Quality Control Log or
otherqualitycontrollog.Thelogshouldalsoincludethelotnumberandexpirationdateofthecontrolsolutionsand
urine test strips.
• Anyqualitycontrolresultthatfallsoutsidetheacceptablecontrolrange,alongwithanycorrectiveactiontorestore
that result to acceptable range, is recorded in the Urinalysis Quality Control Log or other quality control log.
• AnappropriateauthorityoranappointedindividualreviewstheUrinalysis Quality Control Log for completeness and
notes any trends that could indicate potential problems. Such trends include gradual drifting of values, sudden
shifts in urine control values while using the same lot of strips, and operator performance. The log is reviewed
___________ (when? weekly, monthly, quarterly).
• Anaudittrailshouldexisttolinkthecontroltestvaluestothepatienttestvaluesperformedthatdaybylistingthe
appropriate Urisys 1100
®
Urine Analyzer, Chemstrip urine test strip and control solution lot number used.
• Insertyourfacility’spolicyforwhoisresponsiblefortheperformanceanddocumentationofqualitycontrol.
_________________________________________________________________________________________________
_________________________________________________________________________________________________
_________________________________________________________________________________________________
_________________________________________________________________________________________________
_________________________________________________________________________________________________
_________________________________________________________________________________________________
16
J. Limitations of the Procedure
PleasereferencethemostrecentChemstrip urine test strip package insert for modifications in the limitations of proce-
dure.
• Specific Gravity: The chemical principle of the test may cause slightly different results compared with other spe-
cificgravitymethodswhenelevatedamountsofcertainconstituentsarepresent.Glucoseandureaconcentrations
greater than 1% may cause a low specific gravity reading relative to other methods. In the presence of moderate
amounts of protein (100-500 mg/dL) or ketoacidosis, readings may be elevated.
• Leukocyte: This test is not affected by erythrocytes in concentrations up to 10,000/µL or by bacteria common in
urine.Donotcollectspecimensincontainersthathavebeencleanedwithstrongoxidizingagents.Donotusepre-
servatives.Thedrugscephalexinandgentamicinhavebeenfoundtointerferewiththistest.Nitrofurantoincolors
theurineandthuscausesinterferencewithvisualinterpretationoftheteststrip.Highlevelsofalbumin(>500mg/
dL) in the urine may interfere with the test results.
• Nitrite: Large amounts of ascorbic acid (see under glucose) decrease the sensitivity of the test. False positive read-
ings may be produced by medication that colors the urine red or that turns red in an acid medium (e.g., phenazo-
pyridine).
• Protein: False positive results may be found:
Instronglyalkalineurine(pHof9orhigher).
During therapy with phenazopyridine.
When infusions of polyvinylpyrrolidone (blood substitutes) are administered.
Whenresiduesofdisinfectantscontainingquaternaryammoniumgroupsorchlorohexidinearepresentinthe
urine container.
• Glucose: The effect of ascorbic acid (vitamin C) retained in the urine due to ingestion of vitamin tablets, antibiotics,
or fruit juices has been eliminated at glucose concentrations of 100 mg/dL and above so that false negative read-
ings rarely occur, even at high concentrations of ascorbic acid. False positive readings may be produced by strongly
oxidizingcleaningagentsintheurinecontainer.
• Ketone: Red-orange to red color shades, which are readily distinguished from the colors obtained with ketone bod-
ies, can be produced by phenylketone or phthalein compounds administered for liver and kidney function tests. 2-
Mercaptoethanesulphonatesodium(MESNA)orothersulfhydryl-containingcompoundsmaycausefalsepositive
results.
21
• Urobilinogen:Thetotalabsenceofurobilinogencannotbedetected.Mosturinespecimensgiveaslightpinkreac-
tion. This test gives the same color reaction with urobilinogen as with stercobilinogen; however, the differentiation is
not of diagnostic importance. Urine from patients treated with phenazopyridine may show a false positive reaction.
Nitrite concentrations above 5 mg/dL or formalin concentrations above 200 mg/dL (used as a preservative) may
cause a decrease in color reaction.
17
• Bilirubin: Large amounts of ascorbic acid present in the urine following the ingestion of medication containing vita-
min C or fruit juices lower the sensitivity of the test. In case of doubt, repeat test on urine voided at least 10 hours
after the last administration of vitamin C. Elevated concentrations of nitrite, as in urinary tract infections, may result
in lower bilirubin values. Large amounts of urobilinogen in the urine affect the color change of the bilirubin test, but
not enough to give a positive result. False positive results may be produced by medications that color the urine red
or turn red in an acid medium (e.g., phenazopyridine).
• Blood:Falsenegativereadingsareobtainedwhenformalinisusedtopreservetheurine.Nitriteinexcessof10mg/
dL in the urine (which is rare in urinary tract infections) delays the reaction. False positive results can be produced
byresiduesofstronglyoxidizingcleaningagentsintheurinecontainer.Urinefrommenstruatingfemalesoccasion-
ally yields a positive result. This test has not been found to be affected by the ingestion of reasonable quantities of
ascorbic acid.
18
II. Preventive Maintenance
Policy – Cleaning the Analyzer
• CleananddisinfecttheexterioroftheUrisys1100
®
Urine Analyzer by using a cloth dampened with a suitable disin-
fectant such as 10% bleach or isopropyl alcohol. Be sure that no liquid enters the analyzer.
• Cleantheteststriptrayeveryday.
Procedure
1. Turn off the power.
2. Remove the test strip tray.
3. Rinse the tray under running water.
4. Remove any crystalline deposits, especially those contaminating the retaining bar mechanism, with a soft brush.
5. Wipe down the test strip tray with isopropyl alcohol or 10% bleach.
6. Dry with a soft, lint-free cloth.
7. Install the test strip tray by holding the tray opposite the side with the gray reference pad and inserting the tray into
the slot below the function keys. Do not touch the gray reference pad.
8. Switch on the instrument. During the self-check process, the system will verify that the reference pad is in suitable
condition for reading test strips. If not, an error message will be displayed.
Policy – Replacing Printer Paper
Replace the printer paper the day it runs out, following the instructions in Section 4.2 of the Urisys 1100
®
Operator’s
Manual.Thiswillensureapaperprint-outofcalibrationresultswillbeavailablethenextdayifcalibrationisperformed
upon instrument start up.
19
Policy – Documentation and Replacement
• Anymaintenance,calibrationorreplacementisdocumentedonthePreventive Maintenance Log.
• Replaceablepartsareavailable.TheycanbeorderedfromyourlocalRocheDiagnosticsdistributororareavailable
from __________________(location in your facility).
Replaceable parts for the Urisys 1100
®
Analyzer are:
Replaceable Part Catalog Number
Teststriptray-TypeCforChemstrip10MD
Teststriptray–TypeNforChemstrip5OB/Chemstrip7
3666735
3666913
AC Adapter 4340647
Powercord 4340612
Reference pads 1907131
Thermal paper 3666751
Interface connection cable 1906186
Keyboard 1248693
Barcode reader w/o trigger ???
Barcode reader cable ???
#SeeyourOperator’sManualtodeterminewhichteststriptrayisinyouranalyzer.
• IfstaffareunabletocorrectaproblemwiththeUrisys1100
®
Urine Analyzer, they are instructed to contact
_______________(location in your facility) or call Roche Diagnostics Technical Service Center at 1-800-428-4674.
• TheappointedindividualreviewsthePreventive Maintenance Log for completion __________________(when?
weekly, monthly, quarterly).
• Insertyourpolicyforwhoisresponsibleforperforminganddocumentingpreventivemaintenance.
Name Title
_______________________________________________ _______________________________________________
_______________________________________________ _______________________________________________
_______________________________________________ _______________________________________________
_______________________________________________ _______________________________________________
_______________________________________________ _______________________________________________
20
III. Infection Control Guidelines
Policy
• Becauseofthepotentiallyhazardousnatureofhandlingurineandthepossibilityofitscontaminationwithblood,it
isrecommendedthatdisposablelatexglovesandlaboratorycoatsbeusedwhencollectingspecimensandperform-
ing test procedures. Consult your facility’s infection control policy.
• Disposablelatexglovesandlaboratorycoatsaretoberemovedandhandswashedthoroughlywithsoapandwater
after completing the test procedure and prior to handling equipment not related to the procedure.
• UrinespecimensshouldbehandledatBiosafetyLevel2asrecommendedforanypotentiallyinfectiousmaterialin
theCentersforDiseaseControl/NationalInstitutesofHealthmanual,Biosafety in Microbiological and Biomedical
Laboratories, 1988. Universal precautions may apply if the urine is contaminated with blood or if required by the
infection control policy of your facility.
24
• Ifaspillshouldoccur,disinfecttheareawitha1:10bleachsolution(1partbleachplus9partswater)orfollowthe
infection control procedures of your facility.
21
IV. Operator Certification/Recertification
Policy
• OperatorsreceiveappropriatecertificationtoperformtestswiththeUrisys1100
®
Urine Analyzer according to your
facility’s policy.
• CertificationisdocumentedandmaintainedontheOperator Certification Log and Skills Checklist. A Knowledge Test
is also provided.
• OnlycertifiedoperatorsmayusetheUrisys1100
®
Urine Analyzer.
• Operatorcompetencychecks(recertification)willbeperformedyearly and recorded on the Operator Certification
Log and Operator Competency Log.
Procedure – Operator Competency
The following procedure may be used to certify/recertify your operators and check their competency on the use of the
Urisys 1100
®
Urine Analyzer.
Note: A valid calibration is required before performing the following within-run precision.
Within-Run Precision
1. The following equipment is assembled prior to testing:
Urisys 1100
®
Urine Analyzer
Chemstrip urine test strips
Qualitycontrolmaterial–1,2,or3levels(accordingtoyourfacility’spolicy)preparedaccordingtothemanufac-
turer’s instructions
Disposablelatexgloves(ifrequiredbyfacility)
Laboratory coat (if required by facility)
Absorbent paper
Operator Competency Log
2. Run control 20 times consecutively on the Urisys 1100
®
Urine Analyzer. Refer to Section I.H. Quality Control Proce-
dure for instructions.
3. Record results on the Operator Competency Log.
4. Determine the result that occurs most frequently (mode) for each parameter and record the result.
22
5. Determine the color block variation (plus or minus the number of color blocks from the mode) for each parameter in
the run of 20. Record the result for each parameter.
6. The color block variation (number of color blocks away from the mode) for each chemistry of the 20 analyses should
be no greater than ± one color block. Refer to Section 1.H. Quality Control Procedure if the color block variation
exceedsplusorminusone.
Example1:If20consecutiveglucoseresultsproduce20resultsof100mg/dL,themodeis100mg/dLandthecolor
block variation is 0. This is acceptable performance.
Example2:If20consecutiveglucoseresultsproduce15resultsof100mg/dLand5resultsof250mg/dL,themode
is 100 mg/dL and the color block variation is + one color block. This is acceptable performance.
Example3:If20consecutiveglucoseresultsproducetworesultsof50mg/dL,16resultsof100mg/dLandtwo
results of 250 mg/dL, the mode is 100 mg/dL and the color block variation is + one and - one color
block. This is acceptable performance.
Example4:If20consecutiveglucoseresultsproducetworesultsof“normal”,16resultsof100mg/dLandtwo
results of 250 mg/dL, the mode is 100 mg/dL and the color block variation would be + one and - two
color blocks. This is unacceptable performance.
Control Recovery
1. Record the control package insert acceptable ranges for the Urisys 1100
®
Urine Analyzer on the Operator Compe-
tency Log. The Urisys 1100
®
results should be compared to the visual acceptable ranges if the Urisys 1100
®
Analyzer
is not referenced on your control package insert.
2. Compare the acceptable ranges stated in the package insert to the results obtained from the within-run precision
step 4.
3. The mode results should fall within the acceptable control ranges. Refer to Section I.H. Quality Control Procedure for
instructions if the results fall outside the acceptable control ranges.
Guidelines
Insert your policy for who is responsible for training and documenting operator certification / recertification with the
Urisys 1100
®
Urine Analyzer.
___________________________________________________________________________________________________
___________________________________________________________________________________________________
___________________________________________________________________________________________________
23
References
1. Kolthoff, I. Acid-Base Indicators.McMillanCo.,NewYork,1937,p.171.
2. Free,H.,G.CollinsandA.Free.Triple-testStripforUrinaryGlucose,ProteinandpH,Clin Chem, 1960, 6:352.
3. Griess,P.,NotesonthePaperofWeselskyandBenedikt.SomeAzoCompounds.Ber. Dtsch. Chem. Ges., 1879,
12:426.
4. Weltmann,O.,MethodfortheSimpleDetectionofUrinaryTractInfections.Wien. Med. Wschr, 1922, 72:618.
5. Fuchs,T.andG.Gutensohn.ValueandLimitationoftheNitriteTestinDiagnosisofPyelonephritis.Dtsch. Med. J.,
1967, 10:343.
6. Sörensen,S.TheMeasurementoftheHydrogenIonConcentrationandItsImportanceforEnzymaticProcesses.Bio-
chem. Z., 1909, 21:131.
7. USPatentNo.4013416.
8. Gyure,W.L.ComparisonofSeveralMethodsforSemiquantitativeDeterminationofUrinaryProtein.Clin. Chem, 1977,
23:876.
9. Keston, A. Abstracts of papers presented at the 129th meeting of the American Chemical Society, Dallas, April, 1956,
p. 31c.
10.Comer,J.SemiquantitativeSpecificTestPaperforGlucoseinUrine.Anal. Chem., 1956, 28:1748.
11.Legal,E.ANewAcetoneReactionanditsApplicabilityfortheExaminationofUrine.Chem. Centr., 1883, 14:652.
12.Chertack,M.andJ.Sherrick.EvaluationofANitroprussideDipTestForKetoneBodies.J. A. M. A., 1958, 167:1621.
13. Kutter, D. et al. Usefulness of a New Test Strip for Detecting Urobilinogen in Urine. Dtsch. Med. Wschr., 1973, 98:112.
14. With, T. Bile Pigments, Chemical, Biological and Clinical Aspects.AcademicPress,NewYork,1968,p.492.
15.Cook,M.,etal.TheDetectionofBloodinUrine.Am. J. Med. Tech., 1956, 22:218.
16.Leonards,J.SimpleTestForHematuriaComparedWithEstablishedTests.J. A. M. A., 1962, 179:807.
17.Braun,J.andW.Straube.ANewRapidTestForDiagnosingMicrohematuria,ComparedWithResultsOfMicroscopic
Examination.Dtsch. Med. Wschr., 1975, 100:87.
18.Kutter,D.etal.UsefulnessofaNewTestStripfortheDetectionofErythrocytesandHemoglobininUrine.Dtsch.
Med. Wschr., 1974, 99:2332.
19.Brühl,P.etal.ClinicalExperienceGainedwithaTestStripUsedtoScreenforErythrocytesandHemoglobinuria.
Therapiewoche, 1976, 26:5193.
20.NCCLS.ProtectionofLaboratoryWorkersfromInfectiousDiseaseTransmittedbyBloodandTissue.NCCLS Doc.
M29-P,7,9,May,1985,342-347.
24
21.Csako,G.FalsePositiveResultsforKetonewiththeDrugMesnaandOtherFree-SulfhydrylCompounds.Clin. Chem.,
1987, 33/2:289.
22.Henry,J.B.etal.Clinical Diagnosis and Management by Laboratory Methods,17thed.SaundersPhiladelphia,1984,
394;1441.
23.Czerwinski,A.W.,etal.FurtherEvaluationoftheGriessTesttoDetectSignificantBacteriuria.Amer. J. Obstet.
Gynec., 1971, 110:677.
24.CentersforDiseaseControl/NationalInstitutesofHealth.Biosafety in Microbiological and Biomedical Laboratories,
1988.
25
V. Appendix
Instructions for Collections of a Midstream Urine
Whenexaminingurinespecimens,everyprecautionmustbetakentoavoidcontamination.Therefore,makesureyour
patients follow the instructions below carefully so that reasonable test results are obtained.
Itisimportanttonotethatsomecleansingtowelettesmayusesoapsthatcontainoxidizingagents.Oxidizingagents
mayinterferewithsomeparametersontheurineteststrips.Therefore,itisextremelyimportantthatthoroughrinsing
take place after cleansing the genitals.
Procedure
1. Do not open the specimen container until immediately prior to use.
2.Procedure
Female
• Kneelorsquatoverabedpanorstandastrideatoiletbowl.
• Separatelabiaminor.
• Cleansemeatus(openingofurethra)threetimes.
• Rinseareawithsterile,water-saturatedsponge.Donotdry.
• Voidforcibly,allowinginitialstreamofurinetopassintothelavatory.
• Collectmidstreaminsterilecontainer.
Male
• Cleanseglansthreetimes.
• Rinseareawithsterile,water-saturatedsponge.Donotdry.
• Voidforcibly,allowinginitialstreamofurinetopassintothelavatory.
• Collectmidstreaminsterilecontainer.
26
Infant
• Cleansegenitaliathreetimes.
• Rinseareawithsterile,water-saturatedsponge.Donotdry.
• Supportinginfantinafacedownposition,stimulateurinationbystrokingparaspinalmuscles.
• Ifinfantfailstovoid,placesterilebagovergenitals.
• Ifinfantdoesnotvoidwithin45minutes,re-preptheareaandapplyanewbag.
• Collecturineinasterilecontainer.
• Donotallowthecontainertotouchthebody.
3. Close the container and present it immediately for urinalysis.
27
Routine Urinalysis Controls Compatible with the Urisys 1100
®
Urine Analyzer
Several commercial controls are available. Controls may vary in number of levels or components, necessity for reconsti-
tutionorreadytouse,typeandvolumeofcontainer,andusefulnessasacontrolformicroscopicexamination.
When selecting a control, consider the above-mentioned variables to identify the control most suitable for your facility’s
needs. Controls typically can be ordered through medical/surgical supply distributors or directly from the manufacturer.
Refer to the following table to determine which control best meets the needs of your facility’s quality control require-
ments.
Urine Controls Chart
Brand Name KOVA-Trol
®
Urinalysis Control
KOVA
®
Liqua-Trol™ Urinalysis
Control
The Dipper
Urine Dipstick Control
Liquichek™
Routine Urine Chemistry
Control
Manufacturer HYCOR
Biomedical Inc.
HYCOR
Biomedical Inc.
Quantimetrix
Corporation
BIORAD
Website www.hycorbiomedical.com www.hycorbiomedical.com www.4qc.com www.bio-rad.com
Address 7272 Chapman Ave.
GardenGrove,CA92641
7272 Chapman Ave.
GardenGrove,CA92641
2005ManhattanBeachBou-
levard
Redondo Beach, CA 90278
4000 Alfred Nobel Dr
Hercules,CA94547
Toll-free No. 800-382-2527 800-382-2527 800-624-8380 800-224-6723
PhoneNo. 714-895-9558 714-895-9558 310-536-0006 510-724-7000
FaxNo. 714-933-3222 714-895-6920 800.845.1834 510-741-6373
Product
Information
Requests
http://www.hycorbiomedical.
com/site/inforequest.html
http://www.hycorbiomedical.
com/site/inforequest.html
Distribution Cardinal,Fisher,PSS,
McKesson,Owens&Minor,
HenrySchein
Cardinal,Fisher,PSS,
McKesson,Owens&Minor,
HenrySchein
Direct, online thru website ,
CardinalHealth,Fisher,
Infolab,Labsco,McKesson,
Owens&Minor,PSS
Direct, online catalog on
website,Cardinal,PSS,
Labsco
Number of
Levels
3:
Negative Norm
Low Abnormal
HighAbnormal/
HighAbnormalw/urob.
2:
Normal
Abnormal
2:
Normal
Abnormal
2:
Level 1-
Negative/Norm.
Level 2-
Positive/Abnormal
Preparation Reconstitution Ready to use Ready to use Ready to use
PackageSize 4x15mL
4x60mL
10x60mL
3x15mL 6x15mL 10x10mL
Stability /
Unopened
Untilexp.date
on vial (18mo.)
Untilexp.date
on vial
Untilexp.date
on vial
Untilexp.date
on vial
Stability /
Opened
5-7 days after recon
4 mo. frozen
Untilexp.date
on vial
90 days 2-8°C 7 days 2-8°C
Microscopic
Values
Yes Ye s No No
28
VI. Log Sheets
This section contains the following blank log sheets that may be copied and used for documentation for the
Urisys 1100
®
Urine Analyzer in your facility.
Operator Certification Log for recording the names of trained Urisys 1100
®
operators and tracking yearly competency
checks.
Operator Competency Logfor recording yearly competency checks for each operator, within-run precision data, and
control recovery data.
Reagent Log—forrecordingthedatereceived,lotnumbersandexpirationdatesofChemstripurineteststrips,
Chemstrip Calibration Strips and controls.
Urinalysis Report Form — for recording patient results.
Quality Control Logfor recording quality control information.
Preventive Maintenance Logfor recording the dates on which maintenance was performed.
Temperature Log (2) — for recording Urisys 1100
®
Urine Analyzer operating temperatures and Chemstrip urine test strip
storage temperatures.
Knowledge Testfor written documentation of operator certification.
Skills Checklist — for documenting operator skills while training on the Urisys 1100
®
Urine Analyzer
29
Operator Certification Log
Urisys 1100
®
Urine Analyzer
Date of Initial
Certification
Operator ID No. Competency
Checks Date
Observer /
Reviewer Initials /
Date
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
30
Operator Competency Log
Urisys 1100
®
Urine Analyzer
OperatorName:___________________________________________________________ Date:_____________________
ChemstripCat.No.______________________________ LotNo.:__________________ Exp.Date:_________________
ControlNameandLevel:_________________________ LotNo.:__________________ Exp.Date:_________________
Within-Run Precision
SG pH LEU NIT PRO GLU KET UBG BILI BLD
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
Mode
Color
block
variation
Note: The color block variation should be no greater than +/- one color block from the mode for acceptable perfor-
mance.
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
31
Operator Competency Log
Urisys 1100
®
Urine Analyzer
Control Recovery
Control Package Insert Ranges Urisys 1100
®
Urine Analyzer
Mode Result from
Within Run Precision
SG
pH
LEU
NIT
PRO
GLU
KET
UBG
BILI
BLD
Note: Reference the visual acceptance range if the Urisys 1100 Urine Analyzer is not referenced on your control pack-
age insert.
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
32
Reagent Log
Urisys 1100
®
Urine Analyzer
Chemstrip
Urine Test Strips
Chemstrip Urine Analyzer
Calibration Strips
Controls
Date
Rec’d
No.Vials Lot No. Exp.Date Date
Rec’d
No.Vials Lot No. Exp.Date Date
Rec’d
No.Vials Lot No. Exp.Date
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
33
Urinalysis Report Form
Urisys 1100
®
Urine Analyzer
PatientName:________________________________________________________________________________________
Age:______________________________ M F
Physician’sName:_____________________________________________________________________________________
CollectionDate:__________________ TestDate:_______________________ OperatorsInitials:__________________
Physical Examination
Color: colorless yellow amber other
Appearance: clear hazy cloudy turbid
Chemical Examination
SpecificGravity 1.000 1.005 1.010 1.015 1.020 1.025 1.030
pH 5 6 6.5 7 8 9
Leukocytes neg 25 75 500
Nitrite neg pos (any pink color is considered positive)
Protein neg trace 30 100 500
Glucose norm 50 100 250 >1000
Ketones neg 15 50 150
Urobilinogen norm 1 4 8 >12
Bilirubin neg 1 3 6
Blood neg trace 50 250
Comments:
____________________________________________________________________________________________________
____________________________________________________________________________________________________
AttachthePatientReportprintouthereifrequiredbyyourfacility.
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
34
Quality Control Log
Urisys 1100
®
Urine Analyzer
Name of control solution used by your facility: ______________________________________________________________
ControlLotNo.:__________________________________ Exp.Date:______________
Chemstrip Cat. No.: ___________________________________________________________________________________
ChemstripLotNo.:________________________________ Exp.Date:_____________
Operator’sInitials:________________________________
Date SG pH LEU NIT PRO GLU KET UBG BILI BLD Yes/
No
Initials Corrective
Action
Level 1
Level 2
Level 1
Level 2
Level 1
Level 2
Level 1
Level 2
Level 1
Level 2
Level 1
Level 2
Level 1
Level 2
Level 1
Level 2
Level 1
Level 2
Level 1
Level 2
Level 1
Level 2
Level 1
Level 2
Note: Reference the visual acceptable range if the Urisys 1100
®
Urine Analyzer is not referenced on your control pack-
age insert.
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
35
Preventive Maintenance Log
Urisys 1100
®
Urine Analyzer
Checktheboxifyouhaveperformedthetask.Dateandinitialthetopofthecolumn.
Date
Initials
1. Adequate urine test
strip supply available.
2. Urine test strips within
expirationdate.
3. Adequate supply of
calibration strips.
4. Adequate supply of
controls available.
5. Controls within
expirationdate.
6. Cleanexteriorof
analyzer with 10%
bleach solution.
7. Clean the tray of
analyzer with 10%
bleach solution.
8. Calibrate the analyzer.
Performweekly.
9. Replace printer paper
on the day it runs out.
10.Operator’smanualand
test strip package
insert available.
Refer to the Urisys 1100
®
UrineAnalyzerOperator’sManualforspecificproceduralinstructions.
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
36
Temperature Log
Urisys 1100
®
Urine Analyzer
Location: ______________________________________________________
Year: _________________________________________________________
Acceptable Range (°C): __________________________________________
Jan Feb Mar Apr May June July Aug Sept Oct Nov Dec
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
37
Temperature Log
Chemstrip Urine Test Strips
Cat. No. _______________________________________________________
Location: ______________________________________________________
Year: _________________________________________________________
Acceptable Range (°C): __________________________________________
Jan Feb Mar Apr May June July Aug Sept Oct Nov Dec
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
38
Knowledge Test
Urisys 1100
®
Urine Analyzer
Name: ___________________________________________ ID No.: ___________________________________________
Date: ____________________________________________ Unit: _____________________________________________
Mark“T”ifthestatementistrueand“F”ifthestatementisfalse.
_____1. Youmustchecktheexpirationdatingonthevialeachtimeyouperformapatientorqualitycontroltest.
_____ 2. Dip the Chemstrip urine test strip briefly (no longer than 1 second) in the urine sample.
_____3. Strongoxidizingcleaningagentsinterferewiththebloodpatchresults.
_____ 4. Correct placement of the Chemstrip urine test strip in the tray of the analyzer is necessary to obtain accurate
results.
_____ 5. It is necessary to bring refrigerated samples to room temperature before testing.
_____ 6. The sample should be in sufficient volume to allow all test pads to be completely wetted.
_____7. RunningthesideoftheChemstripurineteststripalongtherimofthecontainerremovesexcesssample.
_____ 8. If you do not obtain a successful calibration, the analyzer will still process samples.
_____ 9. A positive nitrite test is not normal in fresh, clean-catch midstream urine specimens.
_____ 10. Urine residue on the tray of the analyzer may result in inaccurate results.
_____11.Greenspotsonthebloodpatchareindicativeofunlysederythrocytes.
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
39
Fill in the blanks.
12. The specimen should be tested within ___________ hour(s) of collection, otherwise, refrigerate it in a closed
container.
13. Replace the _____________________ to avoid deterioration of the test strips.
14. WhenusingChemstrip10MDTestStrips,theUrisys1100
®
Urine Analyzer must be calibrated every ________
days.
15. Reasonable amounts of ascorbic acid do not affect the ________ test.
16. The specimen must be thoroughly ________________________, as red cells sink to the bottom and could cause
false positive results.
17. Rinse the test strip tray with a suitable disinfectant ________ (when?).
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
40
Answer Key
1. T
2. T
3. T
4. T
5. T
6. T
7. T
8. F
9. T
10. T
11. T
12. one
13. vial cap
14. seven
15. blood or glucose
16. mixed
17. daily
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
41
Skills Checklist
Urisys 1100™ Urine Analyzer
The trainer should check each activity as it is demonstrated. All activities must be completed before the trainee is certi-
fied to use the Urisys 1100
®
Urine Analyzer.
Specimen Collection
1. User assembles specimen collection supplies
q Specimen container
q Cleansing towelette
q Water-saturated sponge
2. User reviews specimen collection
q States patient preparation steps
q Explainsspecimenstorageandstability
q Describes clean-catch, midstream urine technique
Testing Procedure
1. User assembles testing supplies
q Urisys 1100
®
Urine Analyzer
q Chemstrip Urine Test Strips
q Chemstrip Urine Analyzer Calibration Strips
q Patientspecimen
q Disposablelatexglovesandlaboratorycoat
q Absorbent paper
2. User preparation
q Checksexpirationdateonteststripvial
q Putsondisposablelatexglovesandlaboratorycoat
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
42
3. Sample preparation
q States if refrigerated, bring sample to room temperature
q Mixessamplebyswirlingcappedcontainerseveraltimes
4. Calibration (when using Chemstrip 10MD or Chemstrip 10UA Test Strips)
q Removes one Chemstrip urine analyzer calibration strip
q Does not touch elevated gray areas on strip
q Placescalibrationstrip(padsideup)ontotheteststriptray
q PressesSTARTbutton
q Documents calibration on the Preventive Maintenance Log
5. Testing
q Removes one Chemstrip urine test strip
q Replaces vial cap immediately
q Ensures analyzer test strip tray is clean
q Briefly dips test strip vertically into specimen
q Drawsstripedgealongrimofcontainertoremoveexcessurine
q Touches strip to absorbent paper
q Placestheteststrip(padsideup)ontotheteststriptray
q The handle end of the strip should be held by the clip at the insertion slot
q Obtainsresultsfromtheprintout
q Removes test strip
q Wipes tray with dry lint-free cloth. If there is any residue on the tray, follows proper maintenance as outlined in
the Urisys 1100
®
Urine Analyzer Operator’s Manual.
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
43
6. Results
q Documents results as appropriate for your facility
7. Qualitycontrol
q StateswhentoperformQCtests
q Performstestlikeroutineurine.
q Documents results
q Reviews procedure if the test results fall outside the acceptable range
q If problem persists, calls Roche Diagnostics Technical Service Center at 1-800-428-4674
8. Infection control
q Properlydisposesofspecimen,usedteststripsandcontainer
q Cleans and decontaminates the work area according to infection control policy
q Removesanddisposesofdisposablelatexgloves
____________________________________________________________________________________________________
OperatorSignature Date
____________________________________________________________________________________________________
Trainer Signature Date
Contact your Roche Diagnostics Account Manager for training certificates.
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
44
VII. Regulatory Requirements
©2008 Roche Diagnostics. All rights reserved.
COBAS,Urisys1100
®
andCHEMSTRIParetrademarksofRoche.
45